Full Spectrum Xenon Lamps
Xenon arc lamps give the best reproduction of full spectrum natural sunlight. Q-Lab's lamps are air-cooled (not water-cooled), which makes them more economical, highly efficient and very low maintenance. Standard lamps last for 1500 hours and the Xe-2 uses one lamp per machine. Enhanced lamps last for 3000 hours at normal irradiance and 1000 hours at high irradiance.
Optical Filters
The Q-SUN Xe-2 model’s optical filter lanterns are designed to produce the spectra specified in many global weathering and lightfastness test standards. These standards apply to a variety of industries for outdoor applications, as well as interior applications where sunlight may enter through a window. The lanterns consist of an outer borosilicate or quartz glass cylinder and two sets of 7 inner filters, arranged in a two-tier heptagon.
Temperature Monitoring & Control
Control of temperature is important because it influences the rate of degradation. Specimen exposure temperature is precisely controlled in the Q-SUN Xe-2 tester using black panel and chamber air temperature sensors. The black panel temperature can be set at any point between 35ºC and 105ºC, depending upon whether the black panel is insulated or uninsulated, irradiance level, filter type, lamp age, chamber air temperature and ambient room temperature. See LX-5046 – Q-SUN Specification Bulletin.
Large Specimen Capacity
The Q-SUN Xe-2 tester has a large specimen capacity, supporting 31 specimens of 46 mm x 122 mm each. Its vertically mounted specimen holders are very easy to install and remove. See our specimen mounting page or Accessories Tab for more information.
SOLAR EYE Irradiance Control
The SOLAR EYE Irradiance Control System constantly monitors and controls lamp output to assure precise light exposure and to maximize repeatability and reproducibility of test results. Irradiance control is available at 340nm or 300-400nm TUV (Total UV). See LX-5046 – Q-SUN Specification Bulletin.
Humidity Control
Relative humidity (RH) control is standard in the Q-SUN Xe-2 tester. It allows the user to simultaneously control and display RH, black panel temperature, and chamber air temperature. Note that purified water is required for proper operation of the RH system in the Xe-2 tester. See LX-5046 – Q-SUN Specification Bulletin.
Optional Water Spray
Outdoor moisture attack is simulated via an optional pure water spray in the Xe-2 tester. A spray nozzle located in the top of the chamber can be programmed to operate during either the dark or light cycle. Note that purified water is required for proper operation of the water spray and RH features in the Xe-2 tester. See LX-5046 – Q-SUN Specification Bulletin.
Optional Back Spray (Model Xe-2-BS)
Back spray is available for the Xe-2 tester and is required by some SAE test methods; it allows water to be sprayed on both the front and back of specimens simultaneously. Note that purified water is required for proper operation of the Xe-2 with back spray.
Controls & Optional Software
The Dual Touchscreen Display is designed to be both functional and easy to use, the Q-SUN Xe-2 controller can be programmed in 17 user-selectable languages (English, Spanish, French, German, Italian, Japanese, Chinese, Korean, Czech, Dutch, Polish, Portuguese, Russian, Swedish, Thai, Turkish, Vietnamese) and includes complete self-diagnostic error checking. The controller constantly monitors the status and performance of all systems. It also displays simple warning messages, routine service reminders or performs safety shutdown, as needed.
An external USB port on every Q-SUN Xe-2 tester allows users to perform software upgrades quickly to address key performance issues. For quality systems that require documented proof of test conditions, this USB port can also be used to download tester performance history. Data from the USB export can be emailed directly to Q-Lab's technical support desk for expert troubleshooting and diagnostics.